Regulatory records

FDA, NICE and European Union records of the attention tests and ADHD-related devices named on this site, in one table, with record type, number, decision date and the intended use on the record. · Last updated

This page brings together, in one table, the official records of the attention tests and ADHD-related (attention-deficit/hyperactivity disorder) devices named on this site. Other pages do not repeat record numbers or dates; they link here. The table covers records of the US Food and Drug Administration (FDA), England's National Institute for Health and Care Excellence (NICE) and the European Union (as of October 2026).

Important

A record does not mean that a product makes a diagnosis on its own. Every diagnosis-related record in the table describes the product as an aid: a qualified health professional interprets the result alongside a clinical assessment.

Records (as of October 2026)

The "Intended use on the record" column summarises the use statement on each record (the FDA's Indications for Use) or the recommendation in the guidance. Dates are decision or publication dates.

Product Maker Authority Record type Identifier Decision date Intended use on the record Record
QbTest Qbtech AB FDA 510(k) K040894 2004-06-22 Objective measurement of hyperactivity, impulsivity and inattention to aid in the clinical assessment of ADHD. Results are to be interpreted only by qualified professionals. FDA
QbTest Qbtech AB FDA 510(k) K122149 2012-10-17 Aid in the clinical assessment of ADHD (QbTest version 3.5). FDA
QbTest Qbtech AB FDA 510(k) K133382 2014-03-24 Aid in the clinical assessment of ADHD and in the evaluation of treatment interventions in people with ADHD. FDA
QbCheck Qbtech AB FDA 510(k) K143468 2016-03-22 The same use as QbTest's 2014 record: aid in the clinical assessment of ADHD and in the evaluation of treatment interventions. FDA
QbTest Qbtech AB NICE Diagnostics guidance DG60 (now HTG729) 2024-10-21 Recommended as an option to help diagnose ADHD at ages 6 to 17, only with a standard clinical assessment by a healthcare professional. For diagnosis at 18 and over, and for monitoring treatment response at any age, only in research. NICE
QbCheck Qbtech AB NICE Diagnostics guidance DG60 (now HTG729) 2024-10-21 Only in research, for diagnosis and for monitoring treatment response. NICE
QbTest, QbCheck Qbtech AB European Union Maker's statement — — The maker describes both tests as "MDR certified". No certificate number is given on the page. Qbtech
T.O.V.A. 9.0 The TOVA Company FDA 510(k) K170082 2017-05-17 Objective measurement of attention and inhibitory control to aid in the assessment of attention deficits, including ADHD. FDA
T.O.V.A. The TOVA Company FDA 510(k) K173915 2018-03-22 Visual test: aid in the assessment of attention deficits and in the evaluation of their treatment. Auditory test: aid in the assessment of attention deficits. FDA
T.O.V.A. 9 The TOVA Company European Union EUDAMED device record — — Registered as a class I medical device, status "on the market". EUDAMED
Nesplora Aula (Attention Kids Aula) Nesplora NICE Diagnostics guidance DG60 (now HTG729) 2024-10-21 Only in research, for diagnosis and for monitoring treatment response. NICE
Nesplora Aula Nesplora European Union Maker's statement — — The maker states that its tests meet the essential requirements of Council Directive 93/42/EEC on medical devices and carry the CE mark. Nesplora
NEBA System Lexicor Medical Technology FDA De Novo DEN110019 (submission K112711) 2013-07-15 Uses the EEG theta/beta ratio at electrode Cz, combined with a clinician's evaluation, to aid in the diagnosis of ADHD at ages 6 to 17. For use as confirmatory support for a completed clinical evaluation; not for stand-alone use. FDA
NEBA Compact EEG2R Mobile Headset Neba Health, LLC FDA 510(k) K223628 2023-03-02 Records brain electrical activity (EEG) and transmits it wirelessly for review by a trained professional; gives no diagnostic conclusion. ADHD does not appear in the use statement. FDA
EndeavorRx Akili Interactive Labs FDA De Novo DEN200026 2020-06-15 Digital therapeutic to improve attention function, as measured by a computer-based test (T.O.V.A.), in children aged 8 to 12 with primarily inattentive or combined-type ADHD who have a demonstrated attention issue. May not show benefit in behavioural symptoms such as hyperactivity; used on prescription, as part of a therapeutic programme. FDA
EndeavorRx Akili Interactive Labs FDA 510(k) K231337 2023-12-13 The same use, ages 8 to 17. FDA
EndeavorOTC Akili Interactive Labs FDA 510(k) K233496 2024-06-14 The same kind of use, ages 18 and over. Not a replacement for any treatment; used as part of a therapeutic programme. FDA
EarliPoint System Earlitec Diagnostics FDA 510(k) K213882 2022-06-08 Aid to clinicians in specialised developmental disabilities centres in the diagnosis and assessment of autism spectrum disorder (ASD) in children aged 16 to 30 months. Not for ADHD. FDA
EarliPoint Earlitec Diagnostics FDA 510(k) K230337 2023-06-29 The same use. FDA
EarliPoint System Earlitec Diagnostics FDA 510(k) K243891 2025-03-26 Aid to qualified clinicians in the diagnosis and assessment of ASD in children aged 16 to 30 months who are at risk based on concerns raised by a parent, caregiver or healthcare provider. FDA
EarliPoint Assessment Earlitec Diagnostics FDA 510(k) K253442 2026-03-05 The same use, with the age range extended to 16 to 95 months (7 years). FDA
CNS Vital Signs CNS Vital Signs, LLC FDA Establishment registration and device listing (not a 510(k)) 3006559064 (establishment) — Listed under product code PTY, "Computerized Cognitive Assessment Aid, Exempt". This product code is exempt from 510(k) submission; there is no use statement reviewed by the FDA. openFDA
MOXO d-CPT Neurotech Solutions Ltd. FDA No record found — — Searches for "MOXO" and "Neurotech Solutions" in the 510(k), De Novo and establishment-registration databases returned nothing (2026-10-18). — (queries: method)
MOXO d-CPT Neurotech Solutions Ltd. European Union No maker's statement found — — The maker's pages that were checked mention HIPAA and GDPR compliance and ISO certification; they state no CE mark or medical-device class (2026-10-18). MOXO
Conners CPT 3 Multi-Health Systems (MHS) FDA No record found — — Searches for "Conners" and "Multi-Health" in the 510(k) database returned nothing (2026-10-18). — (queries: method)
IVA-2 BrainTrain FDA No record; maker's statement — — The maker states that IVA-2 is intended to meet the criteria for non-device clinical decision support software and so does not require FDA clearance. BrainTrain

Among the technologies that NICE recommends only for research, the guidance also lists EFSim Test, EFSim Test Web Version and, for adults, Nesplora Attention Adults Aquarium. MOXO, T.O.V.A., Conners CPT 3 and IVA-2 do not appear in its recommendations or its list of tests.

What do the record types mean?

  • 510(k) premarket notification. The maker shows that its device is "substantially equivalent" to a device already legally marketed in the US (the predicate). If the FDA agrees, it sends an order letter that clears the device for sale; this is a clearance, not an approval. (FDA)
  • De Novo classification request. When there is no predicate to compare with, a new type of device is classified through this route. A device classified through De Novo can then serve as a predicate for later 510(k) submissions. (FDA)
  • Establishment registration and device listing. Establishments that make or distribute medical devices for the US market register with the FDA every year and list their products. This is not a review or a clearance. (FDA) According to the FDA's classification record, product code PTY, under which CNS Vital Signs is listed, is a class II code exempted from 510(k) submission. (openFDA)
  • NICE guidance. A recommendation for England's National Health Service (NHS), not a marketing authorisation. DG60, Digital technologies for assessing ADHD, was published on 2024-10-21. NICE has since renumbered it HTG729; the guidance itself has not changed. (NICE DG60)
  • European Union. Since 2021-05-26, medical devices fall under Regulation (EU) 2017/745 (the MDR), which replaced the earlier Directive 93/42/EEC. (European Commission) In EUDAMED, the EU database of devices, the actor registration, device registration, notified bodies and certificates, and market surveillance modules have been mandatory to use since 2026-05-28; the market surveillance module is for the competent authorities and the European Commission. (European Commission) A maker's own CE or MDR statement is shown in the table separately, as a "maker's statement".

In Turkey, medical devices are registered in the Product Tracking System (Ürün Takip Sistemi, ÜTS); this table does not include ÜTS records. How regulation works in Turkey and elsewhere: Regulation and data protection.

How was this table built?

  • FDA records come from the 510(k) endpoint of openFDA, the FDA's open-data interface; De Novo decisions are in the same endpoint. Each record was compared with the FDA's public record page and with the use statement in the record's summary document.
  • "No record found" describes the state of the database on the day of the search. The openFDA queries that returned nothing: api.fda.gov/device/510k.json?search=device_name:moxo, api.fda.gov/device/510k.json?search=device_name:conners, api.fda.gov/device/510k.json?search=applicant:"multi-health", api.fda.gov/device/registrationlisting.json?search=proprietary_name:moxo.
  • Maker's statement rows report what the maker says on its own pages; no independent record confirms them.

What these records do and do not say about how good a test is: Evaluating a test. The products themselves: MOXO d-CPT, Computerised test systems. What happened to NEBA and EndeavorRx later: Lessons from the past.

Sources

  • FDA. Premarket Notification 510(k). fda.gov
  • FDA. De Novo Classification Request. fda.gov
  • FDA. Device Registration and Listing. fda.gov
  • openFDA. 510(k) endpoint, example query: Qbtech records
  • NICE (2024). Digital technologies for assessing attention deficit hyperactivity disorder (ADHD), DG60 / HTG729. nice.org.uk
  • European Commission. Medical Devices – New regulations: Overview. health.ec.europa.eu
  • European Commission. EUDAMED: Overview. health.ec.europa.eu
  • Each row's own record link is in its last column.